dc.description.abstractSensitivity and specificity of rapid antigen detection tests (RADTs) for detection of group A hemolytic streptococcus (GAS) vary. The purpose is to present the first SKUP (Scandinavian evaluation of laboratory equipment for point of care testing) evaluations concerning the assessment of the diagnostic performance and user-friendliness of two RADTs for detection of GAS when used under real-life conditions in primary health care. Throat samples were collected in duplicates at primary health care centers (PHCCs) from patients with symptoms of pharyngitis. The performance of QuickVue Dipstick Strep A test (307 samples) and DIAQUICK Strep A Blue Dipstick (348 samples) was evaluated using culture results at a clinical microbiology laboratory as comparison. The user-friendliness was evaluated using a questionnaire. The diagnostic sensitivity was 92% (90% confidence interval (CI) 87–96%) and 72% (90% CI 65–79%), while the diagnostic specificity was 86% (90% CI 81–90%) and 98% (90% CI 96–99%) for QuickVue Dipstick Strep A test and DIAQUICK Strep A Blue Dipstick, respectively. Both RADTs obtained acceptable assessments for user-friendliness and fulfilled SKUP’s quality goal for user-friendliness. The diagnostic sensitivity for QuickVue Dipstick Strep A test and the diagnostic specificity for DIAQUICK Strep A Blue Dipstick in this objective and supplier-independent evaluation were higher compared with previous meta-analyses of RADTs. However, the diagnostic specificity for QuickVue Dipstick Strep A test and the diagnostic sensitivity for DIAQUICK Strep A Blue Dipstick were lower compared with previous meta-analyses of RADTs.en_US
dc.language.isoengen_US
dc.publisherSpringeren_US
dc.rightsNavngivelse 4.0 Internasjonal*
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/deed.no*
dc.titlePerformance and user-friendliness of the rapid antigen detection tests QuickVue Dipstick Strep A test and DIAQUICK Strep A Blue Dipstick for pharyngotonsillitis caused by Streptococcus pyogenes in primary health careen_US
dc.typeJournal articleen_US
dc.typePeer revieweden_US
dc.description.versionpublishedVersionen_US
dc.rights.holderCopyright the authors 2020en_US
cristin.ispublishedtrue
cristin.fulltextoriginal
cristin.qualitycode1
dc.identifier.doi10.1007/s10096-020-04034-z
dc.identifier.cristin1838590
dc.source.journalEuropean Journal of Clinical Microbiology and Infectious Diseasesen_US
dc.source.pagenumber549-558en_US
dc.identifier.citationEuropean Journal of Clinical Microbiology and Infectious Diseases. 2021, 40, 549-558.en_US
dc.source.volume40en_US