Monte Carlo simulations of a scanning system based on a panoramic X-ray tube with a conical anode
Peer reviewed, Journal article
Published version
Åpne
Permanent lenke
https://hdl.handle.net/1956/9846Utgivelsesdato
2014-08-21Metadata
Vis full innførselSamlinger
Originalversjon
https://doi.org/10.1155/2014/847651Sammendrag
Monte Carlo simulations were used to study photon production in a panoramic X-ray tube with a conical tungsten target to determine the optimal characteristics of the target shape and electron beam configuration. Several simulations were performed for accelerating potentials equal to 250 kV, 300 kV, and 500 kV with electron beams of various radii and anode sizes. The angular distribution of the photon intensity was analysed by numerical calculations for an assembly composed of an X-ray tube and an external collimator with a cylindrical hole to simulate a panoramic scanning system with an X-ray pencil beam.